MiniCAT™ Technical Specifications
| MiniCAT™ Scanner |
| System Description |
MiniCAT™ is a compact, upright, volume computed tomography (VCT) imaging system designed for high-resolution, low-radiation imaging of the sinuses, skull base, and temporal bones. |
| System Dimensions |
40" L x 36" W x 67" H
(102 cm x 91 cm x 170 cm) |
| Recommended Room Size |
8' x 10' |
| Weight |
450 lbs |
| Internet Connectivity |
High speed connection is required for service/maintenance |
| Patient Support |
| Overall Dimensions |
28" L x 18” W x 50” H
(71 cm x 46 cm x 127 cm) |
| Weight |
115 lbs (52 kg) |
| Seat Height Adjustment |
15.5” to 42"
(39 cm to 107 cm) |
| Maximum Patient Weight |
350 lbs (159 kg) |
| Front View |
Top View |
 |
 |
| Imaging Properties |
| Slice Thickness |
0.3 mm minimum |
| Reconstruction Time |
<60 sec for 600 frames |
| Field of View, Axial |
12 cm max. |
| Field of View, Cross-sectional |
16.5 cm max. |
| Views |
MPR, Axial, Coronal, Sagittal, Oblique, 3D, TMJ, Side-by-Side |
| Gray Scale Display |
256 levels |
| Display Window |
Adjustable level and width |
| Workstation |
Intel-based PC |
| Compatibility |
DICOM 3.0 |
| Acquisition Protocols |
| |
Tube
Voltage |
Tube
Current |
Pulse
Length |
Number of Frames |
Exposure |
Scan Time |
Default Reconstruction Protocol |
| Sinus 20s (600) |
120 kVp |
7 mA |
11.5 ms |
600 |
48.30 mAs |
20s |
SMOOTH (400x400, 0.4, 0.4 f1) |
| Sinus 10s (300) |
120 kVp |
7 mA |
11.5 ms |
300 |
24.15 mAs |
10s |
SMOOTH (400x400, 0.4, 0.4 f1) |
| Sinus 10s (150) |
120 kVp |
7 mA |
11.5 ms |
150 |
12.08 mAs |
10s |
SMOOTH (400x400, 0.4, 0.4 f1) |
| T Bone 20s (600) |
125 kVp |
7 mA |
14 ms |
600 |
58.80 mAs |
20s |
SHARP (536x536, 0.3, 0.3, f0) |
| Resolution QC |
125 kVp |
7 mA |
14 ms |
600 |
58.80 mAs |
30s |
SHARP (800x800), 0.2, 0.2, f0) |
| Water QC |
120 kVp |
7 mA |
11.5 ms |
600 |
48.30 mAs |
20s |
SMOOTH (400x400, 0.4, 0.4 f1) |
| Reconstruction Protocols |
| |
Coronal Slices |
Sagittal
Slices |
Voxel Width/Depth |
Voxel
Height |
Filter
Type |
| SMOOTH (400x400, 0.4, 0.4 f1) |
400 |
400 |
0.4 |
0.4 |
1 |
| SHARP (536x536, 0.3, 0.3, f0) |
536 |
536 |
0.3 |
0.3 |
0 |
| SHARP (800x800), 0.2, 0.2, f0) |
800 |
800 |
0.2 |
0.2 |
0 |
| Filter Type |
| |
Description |
| 0 |
Provides sharp image detail with high-contrast resolution |
| 1 |
Provides smooth image detail with low noise |
| Installation Requirements |
| Recommended Room Size |
8' x 10' |
| Internet Connectivity |
High speed connection is required for service/maintenance |
Note: Room layout must be such that operator of equipment is able to make visual and
audio contact with patient being scanned while operating the equipment from behind
operator's barrier. |
| Electrical Power Requirements |
| |
Europe |
USA |
| Line Voltage* |
~220-240 V |
~115 V |
| Line Frequency |
50 - 60 Hz |
50 - 60 Hz |
| Line Current |
2.5 A Standb |
5 A Standb |
| Main Circuit Breaker |
5 A Max. |
10 A Max. |
| * Voltage must be held to within +- 10% of specified Line Voltage. |
| Environmental Conditions |
| It is recommended that MiniCAT™ is operated in a climate-controlled environment where the ambient temperature does not exceed a range of 60 to 75 degrees Fahrenheit. An operating environment where this suggested range cannot be regulated may reduce the expected life cycle of the MiniCAT™ system. Changes in temperature may cause the system to require additional calibration. |
| Operating Requirements |
50 to 95 degrees Fahrenheit
(10 to 40 degrees Celsius)
45% to 75% humidity |
| Transportation and Storage Requirements |
32 to 104 degrees Fahrenheit
(0 to 40 degrees Celsius)
30% to 80% humidity |
| Heat Dissipation |
Approx. 500W / 1700 BTU/hr |
| Caution: Allow system temperature to adapt to room temperature before powering up MiniCAT™ scanner hardware or workstation/server. |
| Installation / Relocation |
MiniCAT™ cannot be installed or relocated by the user. Proper and safe setup requires
specialized training and must be performed by a Xoran Customer Service Engineer. Call Xoran 1-800-70-XORAN (800-709-6726) for assistance. |
| IGS Compatibility |
The following tables summarize the compatibility of MiniCAT™ with image guided surgery
systems. |
| GE |
| Platform |
Compatible with Xoran |
| IT3500+, ENTrak+, ConneCTstat+ |
Yes |
| IT3500, ENTrak, ConneCTstat |
No |
| IT3000 |
No |
| IT2000 |
No |
| |
|
| Medtronic / LandmarX / Xomed |
| Platform |
Compatible with Xoran |
| LandmarX Evolution, LandmarX Evolution Plus, LandmarX Element |
Yes |
| LandmarX gen1, LandmarX gen2 |
No |
| Medtronic Stealth, Treon, Ion, Tria |
No |
| Medtronic Treon Plus (w/ ENT software), Tria Plus (w/ ENT software) |
Yes |
| Medtronic Fusion |
Yes |
| |
|
| BrainLAB (Requires PacXfer Version 5.2 or Higher) |
| Platform |
Compatible with Xoran |
| VectorVision |
Yes |
| VectorVision 2 |
Yes |
| VectorVision Compact (for ENT) |
Yes |
| Kolibri |
Yes |
| |
|
| Stryker |
|
| Platform |
Compatible with Xoran |
| Stryker Navibation System (original) |
Yes |
| NavII |
Yes |
| mNAV |
Yes |
|